Servicing
Leak Check
7-16
TVA2020 Instruction Manual
Thermo Fisher Scientific
Use the following procedure to verify no leaks.
1.
Attach a flow meter capable of reading a least 2 LPM to the probe inlet.
2.
From the SETUP MENU, choose 1=Calib >
5=Flow
.
3.
Press “
1
” for the first flow calibration point.
4.
Adjust the pump power with the up and down arrows for a reading of 1
LPM at the probe inlet.
5.
From the CONTROL MENU, press
1=Pump Off’
.
6.
Record the counts associated with the pump off.
7.
From the CONTROL MENU, press
1=Pump On
.
8.
Plug the inlet of the probe. The counts should drop to within ±200 of
the “Pump Off” recorded value.