Introduction
External Influences
Thermo Fisher Scientific
TVA2020 Instruction Manual
1-11
Bluetooth
Wireless communications module certified to Bluetooth® ver2.0
Support for SPP (Serial Port Profile)
Transmit Class 2 (typical range of 6’)
“This device complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received above the radiated susceptibility limits,
including interference that may cause undesired operation.”
This device contains FCC-ID POOWML-C40
GPS
Frequency 1575
MHz
Channels
up to 12 simultaneously
Update rate
1 Hz
This product is intended for use in indoor and outdoor environments as a
portable instrument carried by a user, as specified in
Table 1–2
. The same
environmental conditions also apply to the sample stream being monitored.
Table 1–2.
TVA2020 External Influences
External Influences
Reference Operating
Conditions
Normal Operating
Conditions
Operative Limits
Transportation and
Storage Limits
Ambient temperature
23 ± 2
°
C
-10 to +45
°
C
-10 and +45
°
C
-20 to +60
°
C
-4 and 140
°
F
Ambient pressure
860 to 1060 mbar
70 to 108 kPa
20 to 108 kPa
Relative humidity
50% ± 10%
FID: 20 to 95%
PID: 20 to 70%
noncondensing
15 and 95%
noncondensing
0 to 100%
Radiated Susceptibility
None
3 V/m
80 to 2700 MHz
Conducted susceptibility
Not applicable, battery operated
Conducted emission
Not applicable, battery operated
Radiated emission
See Part 15 FCC/
Bluetooth requirement
ESD Sensitivity
>6000 Volts
Battery charger supply
voltage
120 or 230 ±1% Vac
100 to 240 Vac
100 to 240 Vac
Not applicable
Battery Charger supply
frequency
50/60 Hz
±
0.5 Hz
50/60 Hz
50/60 Hz
Not applicable
External
Influences
Содержание TVA2020
Страница 1: ...TVA2020 Instruction Manual Toxic Vapor Analyzer Part Number 111755 00 29Jul2014...
Страница 4: ...Thermo Fisher Scientific WEEE Compliance...
Страница 16: ...Tables xii TVA2020 Instruction Manual Thermo Fisher Scientific...
Страница 76: ......
Страница 96: ......
Страница 118: ......
Страница 178: ...Text Data Formats Factory B 14 TVA2020 Instruction Manual Thermo Fisher Scientific SUM OF SQUARES 0 934123 END...
Страница 190: ...Response Factors Response Factors C 4 TVA2020 Instruction Manual Thermo Fisher Scientific...