Maintenance & Troubleshooting
Flow Rate Checks
Thermo Fisher Scientific
SOLA II User Guide 10-3
8.
Remove the plug from the EXHAUST bulkhead and reconnect the
EXHAUST vent line.
9.
Adjust the sample and air regulators to the pressures specified in the
application notes shipped with the instrument or as recorded in the
instrument logbook.
10.
Allow the analyzer to warm up until it stabilizes.
Figure 10–1. Three-way valve for measuring flow rates
1.
Obtain a precision flow measurement instrument such as a bubble meter.
2.
Ensure that all pressure regulators are set to the correct pressures as
shown on the calibration data shipped with the instrument or as recorded
in the instrument logbook.
3.
Ensure that the analyzer is stabilized at normal operating temperatures
before proceeding.
4.
according to Chapter 5.
Flow Rate
Checks
Содержание SOLA II
Страница 1: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB Part of Thermo Fisher Scientific...
Страница 2: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB...
Страница 4: ......
Страница 6: ......
Страница 12: ......
Страница 16: ......
Страница 38: ......
Страница 50: ......
Страница 84: ......
Страница 92: ......
Страница 150: ......
Страница 159: ...TCP IP Bridge The Browser Interface Thermo Fisher Scientific SOLA II User Guide D 9 Figure D 17 Sample results...
Страница 160: ......
Страница 168: ......
Страница 186: ...Software Menu Tree H 2 SOLA II User Guide Thermo Fisher Scientific Figure H 2 Diagnostics subgroups...
Страница 188: ......
Страница 194: ......
Страница 198: ......
Страница 200: ...Thermo Fisher Scientific SOLA II User Guide M 2 THIS PAGE INTENTIONALLY BLANK...