Configuration
Calibration
6-18 SOLA II User Guide
Thermo Fisher Scientific
Figure 6–46. Low calibration
If necessary, the Adjust Reading function in the Calibration Setup menu can
be used to correct the calibration by moving the line without changing the
slope (see Figure 6–47). Enabling the
function will cause the software
to perform this correction automatically.
Figure 6–47. Corrected calibration
Several calibration functions can be performed. These are discussed in the
following sections.