Configuration
Calibration
6-24 SOLA II User Guide
Thermo Fisher Scientific
6.
Begin a calibration for the low concentration standard or for zero
background.
Figure 6–54.
7.
Perform a calibration without the purge cycle for the high concentration
standard. See
“Performing a High Calibration”
You can do the same for the low concentration standard by accessing the
Recalibrate Low menu item (not shown). See
Figure 6–55.
8.
Adjust calibration during operation to match lab results through the
Adjust Reading menu item. Enter a concentration, and the reading is
forced to match the value entered. See
Figure 6–56.