![Thermo Scientific SOLA II Скачать руководство пользователя страница 105](http://html1.mh-extra.com/html/thermo-scientific/sola-ii/sola-ii_user-manual_1105272105.webp)
Maintenance & Troubleshooting
Replacing the Pyrolyzer Heater
Thermo Fisher Scientific
SOLA II User Guide 10-11
Figure 10–12. Heater wires inserted into terminal block
18.
Inspect the O-ring around the open end of the pyrolyzer housing. If
necessary, replace it and apply a lubricant compound, such as Dow
Corning® 111, around it (Figure 10-13).
Figure 10–13. Pyrolyzer housing O-ring
19.
Gently slide the lid into the housing.
Note
The pyrolyzer tube, insulation, and Kaowool insulation should still be
attached to the lid. Slide the lid back in carefully to avoid damaging the
pyrolyzer tube.
σ
20.
Use the six screws and washers removed earlier to secure the lid to the
housing.
21.
With an air gun placed at the end of one fitting, blow out any dust or
debris that may have dropped inside the pyrolyzer quartz or glass tube
(Figure 10-13).
Содержание SOLA II
Страница 1: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB Part of Thermo Fisher Scientific...
Страница 2: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB...
Страница 4: ......
Страница 6: ......
Страница 12: ......
Страница 16: ......
Страница 38: ......
Страница 50: ......
Страница 84: ......
Страница 92: ......
Страница 150: ......
Страница 159: ...TCP IP Bridge The Browser Interface Thermo Fisher Scientific SOLA II User Guide D 9 Figure D 17 Sample results...
Страница 160: ......
Страница 168: ......
Страница 186: ...Software Menu Tree H 2 SOLA II User Guide Thermo Fisher Scientific Figure H 2 Diagnostics subgroups...
Страница 188: ......
Страница 194: ......
Страница 198: ......
Страница 200: ...Thermo Fisher Scientific SOLA II User Guide M 2 THIS PAGE INTENTIONALLY BLANK...