Diagnostics
Input & Output Test Menu
8-2 SOLA II User Guide
Thermo Fisher Scientific
Table 8–1.
Output
Function
Malfunction
Malfunction relay (fail safe).
Off Line
Denotes calibration cycle or suspended analysis (fail safe).
H Alarm
High alarm relay.
HH Alarm
High high alarm relay.
Stream Selected
Indicates the active stream.
Range Selected
RLY 7
Purge fail (fail safe).
RLY 8 Spare
Stream 1 4–20 mA
Press Enter to change the 4–20 mA output for stream 1 to 4, 8, 12,
16, or 20 mA.
Stream 2 4–20 mA
Press Enter to change the 4–20 mA output for stream 2 to 4, 8, 12,
16, or 20 mA.
Sample Solenoid
Controls the sample diverter.
Stream 1 Solenoid
Controls the valve for stream 1.
Stream 2 Solenoid
Controls the valve for stream 2.
Inject Solenoid
Controls the injection valve.
Cal Solenoid
Controls the calibration/stream valve.
Solenoid 6
Controls high/low flow
Solenoids 7–8
See note below
Bench CNTRL ##
Reserved for factory test only; the 2 pound signs (##) indicate the
control output.
Note
Solenoids 7 and 8 are reserved for future applications and are not
installed in the standard unit.
σ
View MUX Analogs
Содержание SOLA II
Страница 1: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB Part of Thermo Fisher Scientific...
Страница 2: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB...
Страница 4: ......
Страница 6: ......
Страница 12: ......
Страница 16: ......
Страница 38: ......
Страница 50: ......
Страница 84: ......
Страница 92: ......
Страница 150: ......
Страница 159: ...TCP IP Bridge The Browser Interface Thermo Fisher Scientific SOLA II User Guide D 9 Figure D 17 Sample results...
Страница 160: ......
Страница 168: ......
Страница 186: ...Software Menu Tree H 2 SOLA II User Guide Thermo Fisher Scientific Figure H 2 Diagnostics subgroups...
Страница 188: ......
Страница 194: ......
Страница 198: ......
Страница 200: ...Thermo Fisher Scientific SOLA II User Guide M 2 THIS PAGE INTENTIONALLY BLANK...