![Thermo Scientific SOLA II Скачать руководство пользователя страница 77](http://html1.mh-extra.com/html/thermo-scientific/sola-ii/sola-ii_user-manual_1105272077.webp)
Configuration
Calibration
Thermo Fisher Scientific
SOLA II User Guide 6-23
3.
Enter the concentration of the low calibration standard. Note that if using
only one calibration standard and no injections to calibrate the
background, enter zero. See
“Performing a Low Calibration at Zero”
more.
Figure 6–51.
4.
Set the number of readings the analyzer should average during calibration
(1–60 readings).
Figure 6–52.
5.
Begin a calibration for the high concentration standard. See
Figure 6–53.