Version 1.20 Modbus Registers
Status Inputs (1)0001 to (1)0025
Thermo Fisher Scientific
SOLA II User Guide E-3
Table E–3.
Status Input
Number
Description
10001
Low Alarm
10002
High Alarm
10003
I/O Board Timeout
10004
Sample Low Flow Alarm
10005
Lamp Voltage Alarm
10006
Pyrolyzer/Oven temperature Fault
10007
Autocal Error
10008
Purge Fault
10009
Stream Selected: 0 = Stream 1, 1 = Stream 2
10010
Valve Fault
10011
Purging
10012
Calibrating
10013
Suspended
10014
Autocal Performed Clear after read
10015
New Cal Factor Clear after read
10016
Chamber Flow Alarm
10017
Grab Sample Active
10018
Grab Sample Ended
10019
Reserved
10020
Reserved
10021
Reserved
10022
Reserved
10023
Reserved
10024
Reserved
10025
Reserved
Note: Stream 1 = Reg 30001/32768 * Reg 30003
Stream 2 = Reg 30002/32768 * Reg 30004
Status Inputs
(1)0001 to (1)0025
Содержание SOLA II
Страница 1: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB Part of Thermo Fisher Scientific...
Страница 2: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB...
Страница 4: ......
Страница 6: ......
Страница 12: ......
Страница 16: ......
Страница 38: ......
Страница 50: ......
Страница 84: ......
Страница 92: ......
Страница 150: ......
Страница 159: ...TCP IP Bridge The Browser Interface Thermo Fisher Scientific SOLA II User Guide D 9 Figure D 17 Sample results...
Страница 160: ......
Страница 168: ......
Страница 186: ...Software Menu Tree H 2 SOLA II User Guide Thermo Fisher Scientific Figure H 2 Diagnostics subgroups...
Страница 188: ......
Страница 194: ......
Страница 198: ......
Страница 200: ...Thermo Fisher Scientific SOLA II User Guide M 2 THIS PAGE INTENTIONALLY BLANK...