Watlow EZ-ZONE® PM6 Series
Software Configuration
Thermo Fisher Scientific
SOLA II User Guide A-13
Parameter
Default
Comment
Set Point Open Limit High
100.0%
The maximum value of the open
loop set point range.
Table A–19.
Oven settings: Setup Page > Output 1
Parameter
Default
Comment
Display
Software
Display
Software
Output Function
Heat
Output Function Instance
1
Output Control
Fixed Time Base
Output Time Base
0.5
Output Low Power Scale
0%
The power output will never be
less than the value specified and
will represent the value at which
output scaling begins.
Output High Power Scale
70%
The power output will never be
greater than the value specified
and will represent the value at
which output scaling stops.
Table A–20.
Oven settings: Setup Page > Output 2
Parameter
Default
Comment
Display
Software
Display
Software
Output Function
Alarm
Output Function Instance
1
Table A–21.
Oven settings: Setup Page > Alarm 1
Parameter
Default
Comment
Display
Software
Display
Software
Alarm Type
Deviation Alarm
Alarm Hysteresis
2°C
Alarm Logic
Open On Alarm
The output condition during an
alarm state.
Alarm Sides
Both
Both high and low sides can
trigger the alarm.
Содержание SOLA II
Страница 1: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB Part of Thermo Fisher Scientific...
Страница 2: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB...
Страница 4: ......
Страница 6: ......
Страница 12: ......
Страница 16: ......
Страница 38: ......
Страница 50: ......
Страница 84: ......
Страница 92: ......
Страница 150: ......
Страница 159: ...TCP IP Bridge The Browser Interface Thermo Fisher Scientific SOLA II User Guide D 9 Figure D 17 Sample results...
Страница 160: ......
Страница 168: ......
Страница 186: ...Software Menu Tree H 2 SOLA II User Guide Thermo Fisher Scientific Figure H 2 Diagnostics subgroups...
Страница 188: ......
Страница 194: ......
Страница 198: ......
Страница 200: ...Thermo Fisher Scientific SOLA II User Guide M 2 THIS PAGE INTENTIONALLY BLANK...