Maintenance & Troubleshooting
The Pyrolyzer Reaction Tube
10-14 SOLA II User Guide
Thermo Fisher Scientific
Figure 10–16. Pyrolyzer tube inserted into the insulation
5.
Holding the pyrolyzer tube in place, apply the Kaowool insulation over the
top of the insulation (Figure 10-17).
Figure 10–17. Apply the Kaowool insulation
6.
Holding the insulation and pyrolyzer tube, gently attach the housing lid
against the insulation, allowing both ends of the tube to come out of the
lid approximately 1/2”. Attach one of the graphite ferrules and one of the
fittings removed earlier to each of the pyrolyzer tube ends. See Figure 10-
18.
Figure 10–18. Reattach the ferrules and fittings
7.
Gently slide the lid back into the housing and secure it using the screws
and washers removed earlier.