Dual PMT Supply Option
Installation
J-4 SOLA II User Guide
Thermo Fisher Scientific
Viewing parameters in the Diagnostics menu will show the PMT voltage of
the selected supply.
1. View Parameters:
a. Cal A Response Factors
b. Cal B Response Factors
c. PMT Voltage (shows only the active supply voltage)
2. Inputs and Outputs Test > Set Outputs > PMT Supply: Toggle the supply
selection (A or B).
When using the At Line-Grab Sample function, either calibration can be
selected to measure the sample. If the sample concentration range is
unknown, the sample can be measured first on the highest calibration range
and, depending on the concentration, measured on the lower calibration range
to obtain higher accuracy on low concentration samples.
The dual PMT supply option can be retrofitted to existing units in the field.
The parts in the table below are required for the field retrofit.
Table J–1.
P/N
Description
98-1049-0
SOLA II Dual Cal software
97-1649-0
SOLA II Dual PMT Supply kit
At Line-Grab Sample
Installation
Содержание SOLA II
Страница 1: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB Part of Thermo Fisher Scientific...
Страница 2: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB...
Страница 4: ......
Страница 6: ......
Страница 12: ......
Страница 16: ......
Страница 38: ......
Страница 50: ......
Страница 84: ......
Страница 92: ......
Страница 150: ......
Страница 159: ...TCP IP Bridge The Browser Interface Thermo Fisher Scientific SOLA II User Guide D 9 Figure D 17 Sample results...
Страница 160: ......
Страница 168: ......
Страница 186: ...Software Menu Tree H 2 SOLA II User Guide Thermo Fisher Scientific Figure H 2 Diagnostics subgroups...
Страница 188: ......
Страница 194: ......
Страница 198: ......
Страница 200: ...Thermo Fisher Scientific SOLA II User Guide M 2 THIS PAGE INTENTIONALLY BLANK...