Watlow EZ-ZONE® PM6 Series
Software Configuration
Thermo Fisher Scientific
SOLA II User Guide A-11
The default settings for the pyrolyzer are provided in the tables on the
following pages.
Table A–17.
Oven settings: Setup Page > Analog Input 1
Parameter
Default
Comment
Display
Software
Display
Software
Sensor Type
RTD 100 Ohm
TC Linearization
S
For Type S thermocouple
RTD Leads
2
The number of leads on the RTD
wired to this input.
Units
Process
The type of units measured by the
sensor.
Scale Low
0.00
The low scale for process inputs.
Scale High
20.00
The high scale for process inputs.
Range Low
-18.0°C
The output low range.
Range High
5537.0°C
The output high range.
Process Error Enable
Off
Process Error Low Value
0.00
Filter
0.5
For smoothing the process signal.
Input Error Latching
Off
Display Precision
Tenths
Oven Settings
Содержание SOLA II
Страница 1: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB Part of Thermo Fisher Scientific...
Страница 2: ...SOLA II Sulfur Online Analyzer User Guide P N 90 1307 0 Revision AB...
Страница 4: ......
Страница 6: ......
Страница 12: ......
Страница 16: ......
Страница 38: ......
Страница 50: ......
Страница 84: ......
Страница 92: ......
Страница 150: ......
Страница 159: ...TCP IP Bridge The Browser Interface Thermo Fisher Scientific SOLA II User Guide D 9 Figure D 17 Sample results...
Страница 160: ......
Страница 168: ......
Страница 186: ...Software Menu Tree H 2 SOLA II User Guide Thermo Fisher Scientific Figure H 2 Diagnostics subgroups...
Страница 188: ......
Страница 194: ......
Страница 198: ......
Страница 200: ...Thermo Fisher Scientific SOLA II User Guide M 2 THIS PAGE INTENTIONALLY BLANK...