About This Manual
Safety and Equipment Damage Alerts
Thermo Fisher Scientific
PDM3700 Instruction Manual
vii
Alert
Description
WARNING
Whenever the PDM3700 case is opened for maintenance,
you must perform a case leak check after reassembling
the case. The system must pass the case leak check prior
to operating the instrument. Correct the leak and re-test
the instrument if the leak check does not initially pass.
Do not operate the instrument if the system does not
pass the case leak check.
▲
CAUTION
PDM3700 maintenance should only be performed by
trained, authorized personnel.
▲
CAUTION
Refer to Section 1 for a comprehensive list of cautions
and warnings for the battery pack assembly.
▲
CAUTION
Always properly dispose of the PDM3700 battery pack
(recycling is preferable, contact Thermo Fisher Scientific
for further information). Do not dispose of the battery
pack in fire or heat.
▲
Equipment Damage
DO NOT use acetone to clean any parts of the PDM3700.
▲
Equipment Damage
When moving the battery pack to access the wire
connections, be sure to minimize stress on the wires and
connectors.
▲
Equipment Damage
Do not attempt to remove the security (non-Phillips head)
screws on the top or front of the PDM3700.
▲