The displayed mass and mass concentration values will display the last valid
measurement value and will not change when the sampler is placed in an
inverted orientation.
Figure 1–3.
PDM3700 flow path
The PDM3700 is powered by one internal lithium-ion battery pack
assembly. The batteries can be charged daily through the
communication/charge cable and charging unit (
Figure 1–2
). Be sure to
charge the PDM3700 unit in a fresh-air environment.
The lithium-ion battery packs are designed to meet all U.S. Department of
Transportation (DOT) criteria, per: UN Manual of Tests and Criteria Part
III sub-section 38.3, T1, T2, T3, T4, T5, T7.
Various safety features have been included in the design of the lithium-ion
battery packs. However, misusing or mishandling the battery packs can
lead to fluid leakage, heat generation, fire or an explosion.
Battery Safety
Safety Test Criteria
Battery Warnings
Mass Transducer
Pump
Ambient
Pressure
Sensor
Filter Loading
Pressure
Sensor
Flow
Differential
Pressure
Sensor
Pulsation Dampener
Heater
Conductive Tube
Orifice
Air Temperature
& Relative
Humidity
Sensor Block
Ambient
Temperature
Sensor
Cyclone
TEOM