Maintenance and Calibration
Auditing the Tilt Sensor
5-18
PDM3700 Instruction Manual
Thermo Fisher Scientific
Figure 5–25.
Tilt Audit screen
3.
The instrument could take up to 45 minutes to stabilize the TE
frequency.
4.
When stable, the computer will prompt to position the instrument
upright as shown in
Figure 5–26
. Place on a vibration-free table and
support instrument so it does not move during test.
Figure 5–26.
Place instrument in upright position
5.
When prompted, position the instrument on the side (mass transducer)
as shown in
Figure 5–27
. Support instrument so it does not move
during test.