Maintenance and Calibration
Installing the PDM3700 Cover Panel
Thermo Fisher Scientific
PDM3700 Instruction Manual
5-23
Figure 5–31.
PDM3700 with cover panel removed
The cover panel of the PDM3700 may have to be removed to perform
certain maintenance tasks (
Figure 5–31
).
WARNING
Whenever the PDM3700 case is opened for maintenance, you
must perform a case leak check after reassembling the case. The system
must pass the case leak check prior to operating the instrument. Correct the
leak and re-test the instrument if the leak check does not initially pass. Do
not operate the instrument if the system does not pass the case leak check.
▲
1.
Inspect the gasket and sealing surface to ensure they are clean and free
of defects. Replace the gasket, if necessary (Seal Replacement Kit: 59-
009961). Install the gasket.
2.
Install the cover panel onto the PDM3700 case. Ensure that the two
screw holes on the top of the cover panel are aligned with the screw
holes on the top of the PDM3700.
3.
Verify that all portions of the cover panel gasket are seated properly
(
Figure 5–32
).
Installing the
PDM3700 Cover
Panel
To install the PDM3700 cover
panel:
Battery Pack
Foam Spacer
A
B