Basic Operation
Removing a TEOM Filter
Thermo Fisher Scientific
PDM3700 Instruction Manual
2-9
Figure 2–10.
Removing the mass transducer
2.
Carefully insert the lower fork of the filter exchange tool under the
TEOM filter that is in the mass transducer so that the filter disk is
between the lower fork and the upper tab of the filter exchange tool.
The tines of the lower fork should straddle the hub of the filter base
(
Figure 2–11
and
Figure 2–12
).
Figure 2–11.
Filter tool
Figure 2–12.
Removing the TEOM filter with the filter tool