Maintenance and Calibration
Performing a Flow Calibration
Thermo Fisher Scientific
PDM3700 Instruction Manual
5-39
26.
Select the “Yes” button to accept the new flow values or the “No”
button to keep the old values. Press the “Restart” button to perform a
new calibration or the “Cancel” button to exit.
27.
Remove the red orifice from the sample tubing and flow meter tubing.
28.
Install the sample line onto the sample inlet assembly.
Note
Care must be taken when reinstalling the sample tube onto the
sample inlet to prevent cracking, gouging, or otherwise damaging the
tubing.
✳