Section 16- Calibration Check
NOTE: The CF-1 Calibration Fluid is supplied in a single use vial.
The CF-1 must be used within one hour of opening the vial.
Exposure to the environment or transferring of the fluid to another
container may cause a significant change in concentration.
Procedure for Removal of Lint Build-up Under Solenoid
On occasion, the top and bottom pedestals may require
“reconditioning”, which involves rubbing the pedestals aggressively
30-40 times with a lab wipe. Some brands of lab wipes shred during
the process and may result in a build-up of lint under the instrument
solenoid. A significant build-up of lint may alter the absorbance
pathlength, resulting in erroneous measurements. Refer to the image
and follow the instructions below to remove this lint build-up.
1.
Lay the instrument on its side with the source fiber (black
fiber optic cable facing up.
2.
Open the arm of the sampling mechanism.
3.
Using a paperclip or a small screwdriver, manually depress
the solenoid plunger and spray compressed air down the
solenoid plunger hole. Be sure to keep the can of
compressed air upright so as not to spray the propellant into
the instrument
.