Section 3-General Operation
spectrophotometry. Genomic DNA, lambda DNA and viscous
solutions of highly concentrated nucleic acids are common examples
known to the molecular biologist. Proteins are subject to denaturation,
precipitation, and aggregation and therefore may require special
handling to ensure sample homogeneity.
Effect of Evaporation and Solvents
Evaporation of the sample during the measurement cycle usually has
just a minimal effect on absorbance readings and may result in a 1-
2% increase in sample concentration. This can be observed in the
field by measuring the same sample successively over time. Highly
volatile solvents, such as hexane, will likely evaporate before the
measurement can be completed. Less volatile solvents such as
DMSO can be used successfully.
Sample Recovery
One of the advantages of the sample retention system is that samples
can be recovered from the upper and lower measurement pedestals
by extraction with a pipette.
Software Architecture and Features
Main Menu
With the sampling arm in the down position, start the operating
software by selecting the following path:
Start
Æ
Programs
Æ
NanoDrop
Æ
ND-1000 (version)
3-4