![Thermo Scientific NanoDrop 1000 V3.7 Скачать руководство пользователя страница 62](http://html.mh-extra.com/html/thermo-scientific/nanodrop-1000-v3-7/nanodrop-1000-v3-7_user-manual_1105196062.webp)
Section 12- Protein Bradford
Step 2:
Measure
Standards
Up to 5 replicates of
each of up to 7
standards can be
measured.
The software will not
allow measurement
of samples until a
minimum of 1
standard and
references – or 2
standards – are
measured.
Polynomial curve
fitting requires more
standard points.
Step 3:
Measure
Samples
Sample
concentrations can
be calculated by
using linear
interpolation (point-
to-point) between
the two standards
flanking the
unknown sample or
by using polynomial
fitting. [Note: In order
to obtain a
concentration value
(mg/ml) the sample
(unknown) must fall
within the limits of
the standard curve].
Standard Curve Features
Standard curves can be saved and reloaded for reference use by
using the ‘Standard Curve’ pull down menu and choosing the’ Save
as’ or ‘Load Standard Curve’ functions. Selecting the ‘View Standard
Curve’ button allows the user to review the standard curve at any
time. Software versions 3.5.1 and higher include the ability to reload a
12-5