Section 12- Protein Bradford
each assay. Additionally, a standard curve ‘set-up’ may be reloaded.
This feature will recall the respective standard series used in a
previously saved standard curve. Both single and multi-point
standard curve generation is incorporated into the software. A
standard curve can be developed using a reference (Modified
Bradford reagent only – no protein) and a single replicate of one
standard. The multi-point standard curve generator allows a
maximum of five replicates for up to seven different standards. There
is no set order in which standards must be run.
A blank must be measured before the standard curve may be
generated. It is advisable to use water as the blank and use the dye
reagent without any protein added as the “0” or reference sample.
There are three general procedural steps to unknown protein
concentration measurements. The requisite order, including
generating the standard curve, is as follows:
Step 1:
Measure the
‘Reference’
(Bradford reagent –
a ‘zero’ Standard)
Note: The software
will guide the user
with instructions in
the large text box on
the right side of the
screen.
12-4