Section 12- Protein Bradford
Protein standards (BSA) for generating a standard curve may also be
provided by the manufacturer for the Bradford assay kit. Follow the
manufacturer’s protocol for the assay including recommended
incubation times and temperature. Additionally, use the respective
standard (e.g. BSA) and dilutions that cover the analytical range
(mg/ml) of interest. Note: Since the NanoDrop 1000 can measure
higher protein concentrations than traditional cuvette-based
spectrophotometers, you may need to supply your own protein
standards at higher concentrations than provided by the
manufacturer.
Unique Screen Features
View Standard Curve (F8):
selecting this button allows the user to
view the standard curve at any time.
Sample Type:
choices are Reference, Standards 1-7, and Sample.
The software will guide you to measure your reference, then at least
one standard before allowing measurement of samples. Reference
samples refer to the dye reagent without any protein added (ie the “0”
sample).
Replicate #:
counter for tracking replicate number during reference
and standard measurement.
Reset all Standards (F11):
clears all replicates of all standards.
Reset 1 Standard (F12):
clears all replicates of the selected
standard.
Absorbance at 595 nm:
the protein-dye complex’s absorbance at
595 nm for the 1mm pathlength.
Cursor
λ
and Absorbance:
this feature allows the user to adjust the
cursor wavelength and view the corresponding absorbance. The
cursor wavelength can be set by the user. Note: The user-selected
wavelength and absorbance are not utilized in any calculations.
ug/ml:
concentration of the sample (unknown).
Making Bradford Protein Measurements
A standard curve is required every time the Bradford assay is run.
Although curves can be saved and reloaded in the NanoDrop 1000
Spectrophotometer software, it is recommended that the user follow
manufacturers’ guidelines and generate fresh standard curves for
12-3