Section 9- Proteins & Labels
normalized at 340 nm). See the “Concentration Calculation (Beer’s
Law)” Appendices for more details on this calculation.
Baseline Type
This application module has two user-selectable “Baseline Type”
options. The default setting is set to normalize the display spectrum
at 750nm. Alternatively, the 400-750 Slope Baseline Type will
normalize the display at 750 nm and accommodate any linear
baseline offset across the 400 to 750 nm range.
The user may also elect whether or not to use a bichromatic
normalization of the 280 nm absorbance value to the absorbance
value at 340 nm.