Section 14- Cell Cultures
14. Cell Cultures
Using an absorbance spectrophotometer to monitor light scattered by
non-absorbing suspended cells is common practice in life science
laboratories. Such applications, more than any other, accentuate the
differences amongst the optical systems of the numerous
spectrophotometer designs.
Note: The most distinct difference between the NanoDrop 1000
Spectrophotometer “absorbance” values for microbial cell cultures
and those observed using classical cuvette based systems will be
attributable to the shorter pathlength (1 mm vs. 1 cm.) Values may not
be exactly 10 fold different as readings are dependent on both the
optics of a specific spectrophotometer as well as the cell type in
suspension.
The Cell Cultures module displays the sample spectrum from 250 nm
to 700 nm. One cursor is fixed at the frequently used wavelength for
monitoring cell suspensions (600nm) while the second cursor can be
set to the wavelength of interest.
Unique Screen Features
600nm Absorbance:
current value of the absorbance at 600 nm with
the Baseline absorbance subtracted. Note: The actual 1 mm
absorbance is displayed.
14-1