Section 10- Protein BCA
Step 3:
Measure
Samples
Sample
concentrations can be
calculated by using
linear interpolation
(point-to-point)
between the two
standards flanking the
unknown sample or
by using polynomial
fitting. [Note: In order
to obtain a
concentration value
(mg/ml) the sample
(unknown) must fall
within the limits of the
standard curve].
Standard Curve Features
Standard curves can be saved and reloaded for reference use by
using the ‘Standard Curve’ pull down menu and choosing the’ Save
as’ or ‘Load Standard Curve’ functions. Selecting the ‘View Standard
Curve’ button allows the user to review the standard curve at any
time. Software versions 3.5.1 and higher include the ability to reload
a standard curve concentration series using the ‘Load Standard Curve
Set-up’.
Delete Standard Points
The Delete Point button allows the user to delete points by first
selecting the data point to be deleted in the table and then choosing
the ‘Delete Point’ button. If erroneous or non-representative readings
are encountered for a specific standard, all replicates of that standard
are cleared by selecting ‘Delete 1 Standard’. Additionally, all
standards can be deleted at once using the ‘Reset all Standards’
button.
10-5