Section 16- Calibration Check
16. Calibration Check
The Calibration Check is found within the Utilities and Diagnostics
module and is accessed through the Main Menu. It is used to confirm
that both pathlengths are within calibration specifications.
A vial of CF-1 is required to run the calibration check procedure. CF-1
is an aqueous potassium dichromate (K Cr O ) solution for use in
confirming calibration of NanoDrop 1000 Spectrophotometers. It is
roughly ten times more concentrated than other commercially
available K Cr O solutions used to confirm calibration of standard
spectrophotometers and is available through Thermo Fisher Scientific
or your local distributor.
2
2
7
2
2
7
Procedure
1.
Ensure the measurement pedestals are clean and that a 1ul
water sample “beads” up on the lower pedestal.
2.
Remove any lint build-up from around the instrument
solenoid. See the instructions at the end of this section for
removal of lint build-up
3.
Open the NanoDrop 1000 Spectrophotometer Calibration
Check Software and follow the prompts in the Customer
Guidance text box of the software.
4.
Enter the Target Absorbance found on the CF-1 vial as
directed in the image below, Typically the target absorbance
is 0.734; the actual value will depend on the lot of CF-1.
16-1