Section 9- Proteins & Labels
Max Absorbance:
used to rescale the upper limit of the vertical axis.
Sample Type:
The same six sample types (options) listed under
Protein A280 (Section 8) are available for purified Proteins & Labels
analysis and concentration measurement. All of the options can be
viewed by clicking the mouse while it is positioned within the sample
type box. The sample type (color-keyed) can be selected by clicking
on the preferred option or by scrolling through the selections using the
up or down arrow keys located to the left of the sample type box. See
Protein A280 (Section 8) for a detailed description of each sample
type.
λ
3:
user-selected wavelength
Norm Abs @
λ
3:
10mm equivalent normalized absorbance at the
respective wavelength.
Dye 1 (or 2):
user-selected dye
Norm Abs @
λ
1:
normalized 10 mm equivalent absorbance of
selected Dye
uM:
concentration based upon selected Dye’s extinction
coefficient. See the “Concentration Calculation (Beer’s Law)” in
the Appendices for more details on this calculation.
Abs ratio
λ
1/
λ
3:
ratio of the absorbance of Dye 1 to the absorbance
at the user-selected wavelength (
λ
3).
mg/ml:
concentration of proteins in the sample calculated using the
absorbance at 280 nm minus the absorbance at 340 nm (i.e.
9-3