Section 6- MicroArray
normalized at 340 nm) and the nucleic acid analysis constant. See
“Concentration Calculation (Beer’s Law)” in the appendix for more
details on this calculation.
260/280:
ratio of sample absorbance at 260 and 280 nm. The ratio of
absorbance at 260 and 280 nm is used to assess the purity of DNA
and RNA. A ratio of ~1.8 is generally accepted as “pure” for DNA; a
ratio of ~2.0 is generally accepted as “pure” for RNA. If the ratio is
appreciably lower in either case, it may indicate the presence of
protein, phenol or other contaminants that absorb strongly at or near
280 nm. See “260/280 Ratio” in the Troubleshooting section for more
details on factors that can affect this ratio.