2 • Features
Doc. 065289-03 10/12
29
Secondary Pump Head
The secondary pump head delivers eluent to the remainder of the
chromatography system (the injection valve, column, and detector). The
waste valve is located on the front of the secondary pump head (see
To open the waste valve, turn the knob one-quarter to one-half turn
counterclockwise. When the waste valve is in the open position, all pump
flow is directed to waste.
NOTE The waste valve must be open when the pump is
being primed using the Prime button. For detailed
priming instructions, see
.
Figure 2-12. Dionex ICS-1100 Pump Components
Secondary
Pump Head
Primary
Pump Head
Outlet Check
Valve
Inlet Check
Valve
Priming
Valve
Waste Valve
Pressure
Transducer