7. Disconnect the heat exchanger outlet line from the guard column inlet.
8. Use an Allen wrench to remove the three screws from the heat exchanger
metal plate and remove the plate (with the tubing) from the heater.
9. Remove the plastic backing from the new heat exchanger (P/N 059979, with
0.25-mm (0.010-in) ID tubing; P/N 060943, with 0.125-mm (0.005-in) ID
tubing) and press the heat exchanger onto the heater. Replace the screws
removed in
.
10. Connect the new heat exchanger inlet line to port C on the injection valve and
connect the outlet line to the guard column inlet.
11. Reinstall the top metal plate and the heater cover.
5.14 Replacing the Eluent Valve
1. Turn off the power to the Dionex ICS-1100.
2. Open the Dionex ICS-1100 front door.
3. To prevent an eluent leak during the valve replacement procedure, disconnect
the eluent line from the reservoir cap
or
plug the line on the left side of the
valve with a coupler (P/N 039056) and a plug (P/N 037268) after you
disconnect the liquid line from the valve in
4. Disconnect the two liquid
lines connected to the
eluent valve (see
5. Unscrew and remove the
two screws that attach the
valve to the component
panel (see
Save the screws.
Figure 5-18. Eluent Valve
Remove
Disconnect
Содержание Dionex ICS-1100
Страница 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Страница 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Страница 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Страница 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Страница 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Страница 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Страница 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Страница 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...