4 • Troubleshooting
Doc. 065289-03 10/12
71
3. Verify that the
Degas
check box is selected.
4. Check the
Degas
settings. Select either the
Monitor
option or the
Cycle
option. If you select
Cycle
, set the time
On
to 30 seconds and
the time
Off
to 600 seconds. If
Cycle
is already selected, try
increasing the time on or decreasing the time off.
•
End-line filter is dirty or clogged
When new, end-line filters (P/N 045987) are pure white. If the system is in
continuous operation, change the end-line filter weekly, or whenever it
becomes discolored. Replace the filter more often if bacterial buildup is
visible or if the eluent does not contain solvent.
NOTE It is especially important to regularly replace end-line
filters when using aqueous eluents, which may
contaminate the filter with bacteria or algae. The
bacterial buildup may not be visible.
•
Blockages in inlet tubing
Kinked or clogged tubing causes the pump to be “starved” for eluent. Replace
the tubing and fittings (see
•
Dirty check valve
Clean or replace the pump check valve (see
•
Liquid leaks at junction between pump head and pump casting
Use the open-end wrench (P/N 014605) provided in the Dionex ICS-1100
Ship Kit (P/N 057905) to tighten the two acorn nuts that attach the pump head
to the pump housing (see
). Tighten the nuts evenly (12 in-lb
torque).
If the leak persists, replace the piston seal (see
•
Liquid leaks from the seal wash port
Replace the piston seal (see
•
Scratched pump piston
Check the pump pistons for scratches and replace if necessary (see
Содержание Dionex ICS-1100
Страница 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Страница 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Страница 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Страница 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Страница 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Страница 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Страница 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Страница 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...