D • FAQ
D.1
How do I hook up an autosampler?
For instructions on how to connect the Dionex ICS-1100 to an autosampler, refer
to the Dionex ICS-1100 installation instructions. Also refer to the autosampler
operator’s manuals, which are included on the Thermo Scientific Reference
Library DVD (P/N 053891).
D.2
How do I print?
Click the
toolbar button in Chromeleon.
D.3
Why are the retention times moving?
Retention times can shift if the pump flow is erratic or if the column or eluent is
contaminated. See
for pump flow rate troubleshooting. If a
contaminated column is suspected, clean the column as instructed in the column
manual. The column manual is included on the Thermo Scientific Reference
Library DVD (P/N 053891).
D.4
How do I adjust retention times?
Retention times are calculated during calibration. The
Use Recently Detected
Retention Time
parameter in the Chromeleon QNT Editor (
General
tab) can be
used to compensate for some types of retention time drifts; for example,
evaporation of volatile components in pre-mixed solvents or an aging column.
Refer to the Chromeleon Help or user’s manual for details.
D.5
When should I remake standards?
Standards are used only for calibration and should always be made fresh (they
have a lifetime of only one week).
Содержание Dionex ICS-1100
Страница 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Страница 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Страница 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Страница 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Страница 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Страница 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Страница 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Страница 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...