Dionex ICS-1100 Ion Chromatography System
28
Doc. 065289-03 10/12
2.4.2
Eluent Valve
2.4.3
Pump
The Dionex ICS-1100 pump is a microprocessor-based isocratic eluent
delivery system. Its variable speed, dual-piston series design ensures
pulse-free pumping for the most demanding applications.
Primary Pump Head
The primary pump head pumps eluent into the secondary head (see
). The check valves, which prevent reverse flow through the
pump, are located on the bottom (inlet) and top (outlet) of the primary
pump head. The priming valve is on the front of the pump head.
To open the priming valve, turn the knob one-quarter to one-half turn
counterclockwise. When the priming valve is open, liquid can flow into
and out of the primary pump head via the port in the front of the valve.
NOTE The priming valve must be open when the pump is
being primed with a syringe or with isopropyl
alcohol. For detailed priming instructions, see
The eluent valve controls the flow from
the eluent reservoir. The valve opens
automatically when the pump is started
and closes when the pump is turned off.
The valve can also be opened and closed
manually from the Chromeleon Control
panel (see
). This lets you
perform service procedures on pump
components without eluent leaks
occurring.
Figure 2-11. Eluent Valve
Содержание Dionex ICS-1100
Страница 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Страница 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Страница 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Страница 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Страница 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Страница 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Страница 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Страница 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...