Dionex ICS-1100 Ion Chromatography System
86
Doc. 065289-03 10/12
5.1.3
Calibrating the Flow Rate
When to Calibrate
If you run the Dionex ICS-1100 Operational Qualification or Performance
Qualification and it fails.
Items Needed
•
0.076-mm (0.003-in) ID yellow PEEK tubing (P/N 049715) to create
14 ± 2 MPa (2000 ± 300 psi) of backpressure (if needed)
•
High purity ASTM filtered, Type I (18 megohm-cm) deionized water
•
Balance capable of weighing more than 10 g with 0.001 g readability
•
Tared beaker
To Calibrate
1. Open the Dionex ICS-1100 Wellness Panel in Chromeleon (see
2. Verify that there is 14 ± 1.4 MPa (2000 ± 200 psi) of backpressure.
3. On the Chromeleon Wellness Panel under
Pump Flow Rate
Calibration
, select
Reset Cal
.
4. Select
1 mL/min
to set the flow rate and begin pumping deionized
water.
5. Allow the pump to stabilize for 20 minutes.
6. Click
5.00 min
to start the timer. Immediately start collecting water
into the tared beaker.
7. Collect the water for exactly 5.00 minutes.
8. Enter the weight of the water (in grams).
9.
Click
Calibrate
to download the value to the Dionex ICS-1100.
10. To recheck the value, wait at least 15 minutes, and then repeat the
calibration.
11. Click
Log
to record the new calibration value in the Audit Trail.
Содержание Dionex ICS-1100
Страница 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Страница 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Страница 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Страница 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Страница 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Страница 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Страница 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Страница 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...