4 • Troubleshooting
Doc. 065289-03 10/12
67
This error occurs when, in order to maintain the selected current, the Dionex
ICS-1100 is required to apply a higher voltage than the suppressor can
support.
To troubleshoot:
1. Reduce the flow rate.
2. Rehydrate the suppressor. Refer to the suppressor manual for instructions.
Suppressor manuals are included on the Thermo Scientific Reference
Library DVD (P/N 053891).
3. If the error persists, replace the suppressor (see
This error occurs if you turn on the suppressor but the Dionex ICS-1100
cannot establish a connection with the suppressor.
To troubleshoot:
1. Check the suppressor cable connection (see
2. If the error persists, replace the suppressor (see
This error message appears when you click the
Shutdown
button on the
Dionex ICS-1100 Control panel in Chromeleon. The message also appears
anytime you turn off the pump flow while the suppressor is on. The
suppressor is automatically turned off to prevent damage to the suppressor.
To troubleshoot:
•
If the pump stopped unexpectedly, see
troubleshooting information.
SUPPRESSOR OVER POWER
SUPPRESSOR OVER VOLTAGE
SUPPRESSOR STOPPED FOR ZERO FLOW RATE
Содержание Dionex ICS-1100
Страница 10: ...Dionex ICS 1100 Ion Chromatography System viii Doc 065289 03 10 12 ...
Страница 20: ...Dionex ICS 1100 Ion Chromatography System 10 Doc 065289 03 10 12 ...
Страница 46: ...Dionex ICS 1100 Ion Chromatography System 36 Doc 065289 03 10 12 ...
Страница 68: ...Dionex ICS 1100 Ion Chromatography System 58 Doc 065289 03 10 12 ...
Страница 148: ...Dionex ICS 1100 Ion Chromatography System 138 Doc 065289 03 10 12 ...
Страница 152: ...Dionex ICS 1100 Ion Chromatography System 142 Doc 065289 03 10 12 ...
Страница 158: ...Dionex ICS 1100 Ion Chromatography System 148 Doc 065289 03 10 12 ...
Страница 168: ...Dionex ICS 1100 Ion Chromatography System Index 10 Doc 065289 03 10 12 ...