Dionex ICS-1100 Ion Chromatography System
72
Doc. 065289-03 10/12
4.5
Pump Does Not Start
•
No power (front
Power
LED indicator is not lit)
Check that the power cord is plugged in.
Check the main power fuses and replace, if needed (see
•
No communication between Dionex ICS-1100 and Chromeleon
Verify that the USB cable is connected correctly. For connection and setup
information, refer to the Dionex ICS-1100 installation instructions.
•
Pump is turned off
Turn on the pump from the Dionex ICS-1100 Control panel in Chromeleon.
•
Flow rate is set to 0
Set the flow rate from the Dionex ICS-1100 Control panel in Chromeleon.
4.6
No Flow
•
Pump waste or priming valve open (see
Close the valves by turning the knobs clockwise until fingertight.
DO NOT
OVERTIGHTEN! Overtightening may damage the valve and the pump
head.
Figure 4-2. Waste and Priming Valves
Waste Valve
Priming Valve