3 • Operation and Maintenance
Doc. 065289-03 10/12
39
3.2
Turning On the System Power
Press the power switch on the
Dionex ICS-1100 rear panel (see
) to turn on the system
power.
shows the Dionex
ICS-1100 conditions at power-up.
Also turn on the power to the
computer and the autosampler (if
used).
Figure 3-3. Dionex ICS-1100 Rear Panel
Feature
Power-Up Condition
Pump
Off
Injection valve
Load position
Cell
Reading current value
Suppressor
Off*
Cell heater
Set to the last value used. The default when the Dionex
ICS-1100 is turned on for the first time is 35 °C.
Column oven temperature
(Optional)
Set to the last value used. The default when the Dionex
ICS-1100 is turned on for the first time is 30 °C.
* When you start the suppressor, the value used last is restored.
Table 3-2. Dionex ICS-1100 Power-Up Conditions
Power
Switch