DS6 Heated Conductivity Cell Components
The cell front cover provides
CELL IN
and
CELL OUT
fittings for
connecting the cell to the suppressor (see
). The remaining cell
components are mounted behind the component panel. To replace, the cell
remove the screws on the cell front cover and pull the entire cell assembly
out through the component panel. See
for cell replacement
instructions.
2.4.9
ER Controller
The ER Controller (ERC 10) is installed in the RFIC-ER compartment at
the factory. During operation in the RFIC-ER mode, the ERC 10
electronics monitor the volume of electrolysis gases (such as hydrogen
and oxygen) in the eluent stream. If the predefined safety limit is
exceeded, the ERC 10 shuts down the Dionex ICS-1100 pump to prevent
the buildup of gases in the eluent bottle.