Observation
Possible Cause
Action
A bubble was present in the
loading blade when it was
used to apply the PCR
reaction to the chip.
If possible, use the AnalysisSuite
™
Software to filter the low quality data
points, or discard the chip and run the
sample again.
When loading the loading blades:
• If you are using a manual pipette,
pipette to the first stop.
• Decrease your pipetting speed.
Observation
Possible Cause
Action
• The chip leaked during
thermal cycling or
imaging.
• A large bubble was
present in the chip
(insufficient Immersion
Fluid).
• Excess Immersion Fluid
is present on the chip
lid.
If present, remove excess Immersion
Fluid from the chip lid and run the
chip again.
If possible, use the AnalysisSuite
™
Software to filter the low-quality data
points.
To minimize leakage:
• Leave a small bubble when filling
chips with Immersion Fluid.
• When covering chips with
Immersion Fluid, do not get fluid
on the sides of the chip base.
Chapter 6
Troubleshooting
Troubleshooting chip images using the Chip View
6
QuantStudio
™
3D Digital PCR System User Guide
55
Содержание Applied Biosystems QuantStudio
Страница 125: ......
Страница 126: ...thermofisher com support thermofisher com askaquestion thermofisher com 18 March 2020 ...