15.
Open the QuantStudio
™
3D Digital PCR Chip Loader package and place the
loader on the surface where you will load chips. Attach the power cord to the
chip loader and connect it to an outlet.
IMPORTANT!
Do not power on the chip loader until you are ready to use it.
IMPORTANT!
Do not install the chip loader in an area that may damage the
enclosure.
IMPORTANT!
When loading chips, the chip loader must be placed on a level
surface.
The instrument is now ready to use.
Install the ProFlex
™
2 × Flat PCR System
This section includes the essential procedures for preparing the ProFlex
™
2 × Flat
PCR System for use. For complete instructions, see the
ProFlex
™
PCR System User
Guide
(Pub. No. MAN0007697).
Note:
For information on preparing the Dual Flat Block GeneAmp
™
PCR System
1.
Open the package containing the Base Module.
2.
Remove the packing crate from the package, and verify that it contains all of the
parts listed on the shipping manifest.
If you have not received one or more of the parts, contact customer support for
replacement (see “Customer and technical support” on page 123).
Unpack and
place the
ProFlex
™
PCR
System Base
Module
Appendix A
Install the QuantStudio
™
3D Digital PCR System
Install the ProFlex
™
2 × Flat PCR System
A
68
QuantStudio
™
3D Digital PCR System User Guide