2.
When the thermal cycler displays the main menu, confirm that the version
number displayed on the screen is
3.12
or later. If the version number is earlier
than 3.12, contact technical support to upgrade the instrument firmware before
continuing.
IMPORTANT!
The GeneAmp
™
PCR System 9700 must be running firmware
version 3.12 or later to thermal cycle the chips.
08:00 AM
4/25/13
25.0°C
GeneAmp® PCR System 9700
Name:tc001 User:<<pe>>
F1
F3
F4
F5
F2
Run
User
Util
Edit
Create
Version: 3.12
1
1
v3.12 or later
3.
From the Main menu, press
Util
(F4).
08:00 AM
4/25/13
25.0°C
GeneAmp® PCR System 9700
F1
F3
F4
F5
F2
Run
User
Util
Edit
Create
Name:tc001 User:<<pe>>
Version: 3.12
4.
From the Utilities screen, press
Config
(F3).
Utilities
Diag - Instrument diagnostics
TmCalc - Calculates melting temp
Config - Instrument configuration
F1
F3
F4
F5
F2
Diag
Exit
More
Config
TmCalc
5.
From the Instrument Configuration screen, set the instrument time:
a.
Use the Circular Key Pad to select the
Time
field.
b.
Press the
24 Hr
(F2) or
PM (AM)
(F3) soft keys until the format you want for
the current time displays in the Time field.
c.
Use the numeric keys to type in the hours followed by minutes.
d.
Press
Accept
(F1) when your entries are complete.
F1
F3
F4
F5
F2
Accept
Cancel
More
PM
24Hr
Instrument Configuration
Time: 11:30 AM
Date: 01/25/00 M/D/Y
Run Time Printer: Off
Run Time Beep: Off
Note:
CE clears an entry.
6.
From the Configuration screen, set the instrument date:
a.
Use the circular key to select the
Date
field.
Appendix A
Install the QuantStudio
™
3D Digital PCR System
Install the GeneAmp
™
PCR System 9700
A
76
QuantStudio
™
3D Digital PCR System User Guide