7.
Ensure that the loading blade is firmly seated on the loader arm.
8.
Press the dark grey/black loading button to load the chip.
1
1
Loading button
The status light flashes green during the loading sequence, and displays solid
green when finished.
9.
After loading, hold the Immersion Fluid syringe at an angle over the chip surface
without touching the surface, and
slowly
add several drops of fluid directly onto
the chip so that the fluid covers the entire surface. After dispensing, remove any
fluid from the edges of the chip case with a low-lint wipe that has been sprayed
with isopropanol.
IMPORTANT!
·
Expel the fluid slowly to avoid flushing out the reactions in the chip wells.
·
Do not allow the syringe tip to touch the chip surface.
·
Cover the entire surface of the chip, or reactions in the exposed wells may
evaporate.
·
Do
not
fill the entire chip case with fluid. A small amount of fluid running off
the surface of the chip is acceptable.
Correct
Fluid on
edges
Incorrect
Insufficient
fluid
Fluid over
chip
Chapter 3
Prepare the digital PCR reaction and load the chips
Load the chips using the QuantStudio
™
3D Digital PCR Chip Loader
3
QuantStudio
™
3D Digital PCR System User Guide
29