c.
Inspect the instrument for damage caused during transportation. If the
instrument is damaged, note the location and appearance of the damage,
then contact customer support or your service representative for
assistance.
6.
If you intend to connect the QuantStudio
™
3D Instrument to a network, prepare
the physical network connection:
Note:
Connecting the instrument to a network is optional and can be done later.
Connection
Action
Wired
Connect one end of a standard category-6 Ethernet cable to the
RJ-45 port on the back panel of the instrument and the other end
to an open network port.
Wireless
1.
If required for network setup, record the MAC address of the
QuantStudio
™
3D Extended Arm WiFi Adapter, which is
located on the underside of the adapter.
2.
Unplug the instrument.
3.
Turn the instrument over, then use your fingers to carefully
pry open the rear panel.
IMPORTANT!
The rear panel is secured to the instrument
by three pegs that require significant force to decouple. If
necessary, insert a flathead screwdriver into the slot at the
base of the panel to pry it from the instrument chassis.
4.
Uncap the WiFi Adapter and insert it into the internal USB
port
inside
the instrument. When the device is seated in the
USB port, firmly push on the adapter to ensure that it is
plugged in completely.
Note:
The internal USB port is accessed through a small
opening in the instrument chassis.
5.
Reattach the rear panel to the instrument chassis.
Appendix A
Install the QuantStudio
™
3D Digital PCR System
Install the QuantStudio
™
3D Digital PCR Instrument
A
64
QuantStudio
™
3D Digital PCR System User Guide