Page
52
of
76
Mix freq. start
10
–
30 oscillations per minute
Default: 30
Sets the agitating frequency of the tray at the start of donation.
Mix freq. end
10
–
30 oscillations per minute
Default: 30
Sets the agitating frequency of the tray at the end of donation. If this frequency
is different from the frequency set in ‘Mix freq. start’, the
oscillation frequency will
change gradually during donation.
Start agitator
0
–
60 ml
Default: 0
Volume to be collected before the tray starts agitating. When set to 0, the agitator
starts immediately.
Keep mixing
Yes / No
Default: Yes
Yes:
After donation, the tray keeps on agitating until the clamp goes open.
No:
After donation, the agitating is stopped.
Following program settings can be found on screen 4:
Stats. timer
Yes / No
Default: Yes
Yes:
The statistics screen proceeds automatically to the idle screen after 8
seconds.
No:
The user
needs to press the ‘finish’ button on the statistics screen in
order to proceed.
Blood density
1.05
–
1.06g/ml
Default: 1.06
The blood volume which is displayed on T-RAC II is based on the weight of the
blood and the blood density. The blood density can be adjusted here.
Proceed weigh
0
–
300g
Default: 20
When the idle screen is displayed, the screen continues automatically to the next
screen when a bag is placed
in the tray. ‘Proceed weigh’ is
the weight difference
needed to trigger this. 0 disables auto-
proceeding, the user needs to press ‘next’
in order to proceed.
Auto-clamp
Yes / No
Default: Yes
Yes:
When tube is inserted in the clamp before donation, the clamp locks
after pushing the clamp head backwards and then releasing it.
No:
The user needs to press the button
‘
lock
clamp’ on the touch screen
for locking the clamp.