Minimum vol.
1
–
650ml
Default: 1
Minimum volume which can be chosen in the event ‘volume chooser’
or which can
be received from BBIS/TOMEs.
Maximum vol.
2
–
650ml
Default: 500
Maximum volume which can be chosen in the event ‘volume chooser’
or which can
be received from BBIS/TOMEs.
Choose vol.
Fix /
BBIS / Barc
/ Prst / Keyb
Default: Keyb
Fix:
The donation volume of preset 1 is chosen
automatically. No ‘volume
chooser’ event will be displayed before donation.
BBIS:
The donation volume is obtained from BBIS/TOMEs. A volume
confirmation message will be shown on the screen instead of the
volume chooser event.
A network connection with TOMEs is required for this to work,
otherwise the donation will need to be cancelled. Also, special
configuration for bi-directional communication is required, please
refer to the TOMEs manual for more information.
Barc:
The
‘volume chooser’ event is activated. The donation volume can
only be chosen via one of the barcodes ‘Barcode vol.’
Prst:
The ‘volume chooser’ event is activated.
The donation volume can be
selected on the touch screen using one of the preset buttons
, or via
the barcodes ‘Barcode vol.’ if they are defined.
Keyb:
The ‘volume chooser’ event is activated.
The donation volume can be
selected on the touch screen using one of the preset buttons, by using
the keyboard (via the
button)
or via the barcodes ‘Barcode vol.’
if they are defined.
Preset vol. 1
–
Preset vol. 4
0
–
650ml
Default: 400, 450, 500 and 0ml
Here the donation volume presets are defined which can be chosen in the event
‘volume chooser’. When a preset is set to 0, it will not be
shown in the event. The
volumes which can be chosen here, are not limited to the limits set in the settings
‘Minimum vol.’ and ‘Maximum vol.’.
The first preset cannot be set to 0.
Barcode vol. 1
–
Barcode vol. 4
Barcodes (max 25 characters)
A barcode can be assigned to each donation volume preset. When the setting
‘Choose vol.’ is set to Barc, Prst or Keyb, the barcode can be read to select the
corresponding volume.
Following program settings can be found on screen 3:
Low flow limit
1
–
200ml/min
Default: 40ml/min
Minimum flow rate for triggering the low flow rate alarm and the low flow rate pre-
alarm.