AC Power Distribution
Assembly—Z1884
The AC power distribution assembly for Z1884 contains power line monitoring circuitry
which ensures that acceptable electrical voltages are maintained within the system.
Important:
The LEDs are not safety indicators.
Item #
Name
Function
1
Ground
Connects to ground.
2
Switched System Outlets
Eight 120 VAC outlets provide AC distribution to tester.
When line monitor detects faults, it cuts power to these
outlets.
3
Power Input
Connects AC power.
4
CB1
25 A AC Mains circuit breaker to switch AC power on/
off.
5
CB2
5 A user outlets circuit breaker for four user unswitched
outlets.
6
CB3
15 A circuit breaker to turns off switched outlets inside
the tester.
7
Unswitched User Outlets
Four 120 VAC outlets to provide unswitched power.
8
Fuse
Line monitor board protection.
9
LEDs
Green LED indicates AC power is on. It remains on as
long as AC is on. Red LED indicates detection of out-
of-range voltage or an emergency off signal.
120V
5A 50/60HZ
I
O
CB2
I
O
CB3
120V
5A 50/60HZ
T .16A 25V
1
2
3
4
5
6
7
8
9