1-37
Z1800-Series Operator’s Guide
Z1860M
Z1860M Safety Notes—Front, Top and Right Side View
1. Test jack panel
•
When the analog debug LED is on, and the High Voltage Option is installed, high
voltage may be present on the E, F, and G test points.
•
The wrist strap receptacle grounds the antistatic wrist strap. Connect the wrist strap to
one of the receptacles on the tester and wear it at all times when you are working on the
system except when you are working on the AC power control or when you are
otherwise instructed.
2. Operator control pod
•
The Power Control switch turns power to the tester on and off. However, when the
Power Control switch is off, the unswitched outlets are still powered. See the
“Emergency Off Procedures” section for more detailed information.
•
Pressing the Cancel switch stops test execution and returns to the start of the program.
See the “Emergency Off Procedures” section for further details.
3. AC power distribution assembly
3.1.
AC power input
Makes protective ground connection.
Hazardous voltage and current.
3.2. CB2
Mains circuit breaker, 15 A.
3.3. CB1
Circuit breaker for user outlets, 5 A.
3.4.
Wrist strap receptacle
The receptacle is provided primarily for grounding
the antistatic floor mats. You can also connect the antistatic wrist strap to this
receptacle.
3.5. Ground
3.6. User
outlets
Four 120 VAC outlets.
Electrical hazard is present whenever the main AC breaker is on, and AC power
is connected to the power inlet.
3.7. LEDs
The green LED indicates AC power is on. The red LED indicates a
fault.
3.8. Fuse