1-32
SAFETY
9. AC power distribution assembly
9.1. Outlets
120 VAC outlets.
Electrical hazard present whenever Circuit Breaker 3 and the main AC breaker
is on, and AC power is connected to the power inlet.
9.2.
Main AC circuit breaker
Do not connect or disconnect the AC power cord
while the main AC breaker is on; damage to the tester can result. Hazardous
voltage present whenever AC power cord is attached.
9.3. CB2
Circuit breaker for user outlets, 5 A.
9.4.
CB3
Circuit breaker for switched system outlets.
9.5.
Ground
Used to connect your facility ground to the tester ground system.
Z1803 Plus—Z1803-1,
Z1803-2
Z1803 Plus—Front, Top, and Right Side
1. Emergency Manual Off (EMO) switch, optional
Push to activate. Turn to deactivate. Activating this switch has the same effect as
pressing the Power Control switch on the operator control panel. Power is still present
on the line conditioner, AC power control, and unswitched user outlets.
2. Fixture receiver
Warning. Hazardous voltages may be present on fixture receiver during test.
•
Do not remove fixture receiver covers while a program is running. Power may be
applied to the fixture receiver, consequently placing you at risk.
•
Keep fingers away from fixture receiver pins.
•
Remove rings, watches, bracelets, or any loose or conductive apparel.
•
Observe proper ESD precautions.
1
6
2
3
4
5